Forums before death by AOL, social media and spammers... "We can't have nice things"
|    sci.electronics.repair    |    Fixing electronic equipment    |    124,925 messages    |
[   << oldest   |   < older   |   list   |   newer >   |   newest >>   ]
|    Message 124,019 of 124,925    |
|    Peter W. to All    |
|    Re: Failure of a pair of Eppindorf centr    |
|    23 Nov 23 04:33:37    |
      From: peterwieck33@gmail.com              Same age? Close serial numbers?              As to power - did you plug in another piece of equipment, or just check with       some sort of tester? If the neutral is out, you may have "power", but not       usable power.              Suggestions in no particular order:              a) Find the main panel that feeds those circuits and the individual       breaker(s). Cycle them.       b) Move the machines to another known-good receptacle.       c) Check the power cords. If they are not hard-wired, try removing/plugging       back in.       d) Check the lid interlock(s). If they are sticking, the machines will not       start. In the case of some of the machines here, they will not even light up       if the interlock is not either fully open or fully engaged.       e) Replace the fuses anyway. They get tired and may not show as blown, but       may still be open.              Good Luck!              Peter Wieck       Melrose Park, PA              --- SoupGate-Win32 v1.05        * Origin: you cannot sedate... all the things you hate (1:229/2)    |
[   << oldest   |   < older   |   list   |   newer >   |   newest >>   ]
(c) 1994, bbs@darkrealms.ca